Scanned-Fringe, Spatial-Harmonic-Distortion Test For Detector Nonlinearity

Authors

    Authors

    A. J. Varela;G. D. Boreman

    Comments

    Authors: contact us about adding a copy of your work at STARS@ucf.edu

    Abbreviated Journal Title

    Opt. Eng.

    Keywords

    INFRARED TECHNOLOGY; DETECTORS, NONLINEARITY; SPATIAL-HARMONIC-DISTORTION TEST; Optics

    Abstract

    Results are presented for a modification of the spatial-harmonic-distortion test, suitable for nonlinearity measurement of single-element detectors, using a scanned-fringe technique and an electronic spectrum analyzer. Detector nonlinearity of less than 1% was measured at a wavelength of 10.6 mum. Previously, the smallest measured nonlinearity for the spatial-harmonic-distortion test was of the order of 3%, using a stationary-fringe technique and an 8-bit frame grabber.

    Journal Title

    Optical Engineering

    Volume

    33

    Issue/Number

    3

    Publication Date

    1-1-1995

    Document Type

    Article

    Language

    English

    First Page

    721

    Last Page

    724