Title

Exact filtered backprojection reconstruction for dynamic pitch helical cone beam computed tomography

Authors

Authors

A. Katsevich; S. Basu;J. Hsieh

Comments

Authors: contact us about adding a copy of your work at STARS@ucf.edu

Abbreviated Journal Title

Phys. Med. Biol.

Keywords

LONG-OBJECT PROBLEM; INVERSION ALGORITHM; CT; Engineering, Biomedical; Radiology, Nuclear Medicine & Medical Imaging

Abstract

We present an exact filtered backprojection reconstruction formula for helical cone beam computed tomography in which the pitch of the helix varies with time. We prove that the resulting algorithm, which is functionally identical to the constant pitch case, provides exact reconstruction provided that the projection of the helix onto the detector forms convex boundaries and that PI lines are unique. Furthermore, we demonstrate that both of these conditions are satisfied provided the sum of the translational velocity and the derivative of the translational acceleration does not change sign. As a special case, we show that gantry tilt can also be handled by our dynamic pitch formula. Simulation results demonstrate the resulting algorithm.

Journal Title

Physics in Medicine and Biology

Volume

49

Issue/Number

14

Publication Date

1-1-2004

Document Type

Article

Language

English

First Page

3089

Last Page

3103