Authors

D. Mandridis; I. Ozdur; M. Bagnell;P. J. Delfyett

Comments

Authors: contact us about adding a copy of your work at [email protected]

Abbreviated Journal Title

Opt. Express

Keywords

FREQUENCY STABILIZATION; OPTICAL CAVITY; LASER; PRECISION; Optics

Abstract

In this work a narrow linewidth (1 kHz) laser source is used to measure the free spectral range of a fiberized Fabry-Perot etalon with sub-Hz accuracy (10(-8)). A previously demonstrated technique based on the Pound-Drever-Hall error signal is improved in accuracy by the use of a narrow linewidth laser swept in frequency via an acousto-optic modulator, or single sideband generation. The sub-Hz (10(-8)) accuracy attained enables the characterization of both the long-term drift and the polarization dependence of the free spectral range of the fiberized etalon.

Journal Title

Optics Express

Volume

18

Issue/Number

11

Publication Date

1-1-2010

Document Type

Article

Language

English

First Page

11264

Last Page

11269

WOS Identifier

WOS:000278512300040

ISSN

1094-4087

Share

COinS
 

Accessibility Statement

This item was created or digitized prior to April 24, 2027, or is a reproduction of legacy media created before that date. It is preserved in its original, unmodified state specifically for research, reference, or historical recordkeeping. In accordance with the ADA Title II Final Rule, the University Libraries provides accessible versions of archival materials upon request. To request an accommodation for this item, please submit an accessibility request form.