Title
Low Dose Plastic Optical Fibre Radiation Dosimeter For Clinical Dosimetry Applications
Abstract
PMMA optical fibres offer the opportunity to monitor ionising radiation doses online and in real-time. The work presented here investigates the use of these fibres in clinical dosimetry applications. Tests to determine the sensitivity of the PMMA optical fibres at low radiation doses were carried out at the MD Anderson Cancer Center Orlando, Florida. The results demonstrate that PMMA optical fibres exhibit a quantifiable response to low doses (up to 12Gy) of radiation, similar to that received during radiation treatment. ©2009 IEEE.
Publication Date
12-1-2009
Publication Title
Proceedings of IEEE Sensors
Number of Pages
1689-1692
Document Type
Article; Proceedings Paper
Personal Identifier
scopus
DOI Link
https://doi.org/10.1109/ICSENS.2009.5398516
Copyright Status
Unknown
Socpus ID
77951126841 (Scopus)
Source API URL
https://api.elsevier.com/content/abstract/scopus_id/77951126841
STARS Citation
O'Keeffe, S.; Lewis, E.; Santhanam, A.; Winningham, A.; and Rolland, J. P., "Low Dose Plastic Optical Fibre Radiation Dosimeter For Clinical Dosimetry Applications" (2009). Scopus Export 2000s. 11422.