Three-Dimensional Reciprocal Space X-Ray Coherent Scattering Tomography Of Two-Dimensional Object
Keywords
coherent scattering; computed tomography (CT); small-angle x-ray scattering (SAXS)
Abstract
Purpose: X-ray coherent scattering tomography is a powerful tool in discriminating biological tissues and bio-compatible materials. Conventional x-ray scattering tomography framework can only resolve isotropic scattering profile under the assumption that the material is amorphous or in powder form, which is not true especially for biological samples with orientation-dependent structure. Previous tomography schemes based on x-ray coherent scattering failed to preserve the scattering pattern from samples with preferred orientations, or required elaborated data acquisition scheme, which could limit its application in practical settings. Here, we demonstrate a simple imaging modality to preserve the anisotropic scattering signal in three-dimensional reciprocal (momentum transfer) space of a two-dimensional sample layer. Methods: By incorporating detector movement along the direction of x-ray beam, combined with a tomographic data acquisition scheme, we match the five dimensions of the measurements with the five dimensions (three in momentum transfer domain, and two in spatial domain) of the object. We employed a collimated pencil beam of a table-top copper-anode x-ray tube, along with a panel detector to investigate the feasibility of our method. Results: We have demonstrated x-ray coherent scattering tomographic imaging at a spatial resolution ~2 mm and momentum transfer resolution 0.01 Å−1 for the rotation-invariant scattering direction. For any arbitrary, non-rotation-invariant direction, the same spatial and momentum transfer resolution can be achieved based on the spatial information from the rotation-invariant direction. The reconstructed scattering profile of each pixel from the experiment is consistent with the x-ray diffraction profile of each material. The three-dimensional scattering pattern recovered from the measurement reveals the partially ordered molecular structure of Teflon wrap in our sample. Conclusions: We extend the applicability of conventional x-ray coherent scattering tomography to the reconstruction of two-dimensional samples with anisotropic scattering profile by introducing additional degree of freedom on the detector. The presented method has the potential to achieve low-cost, high-specificity material discrimination based on x-ray coherent scattering.
Publication Date
4-1-2018
Publication Title
Medical Physics
Volume
45
Issue
4
Number of Pages
1654-1661
Document Type
Article
Personal Identifier
scopus
DOI Link
https://doi.org/10.1002/mp.12813
Copyright Status
Unknown
Socpus ID
85043571690 (Scopus)
Source API URL
https://api.elsevier.com/content/abstract/scopus_id/85043571690
STARS Citation
Zhu, Zheyuan and Pang, Shuo, "Three-Dimensional Reciprocal Space X-Ray Coherent Scattering Tomography Of Two-Dimensional Object" (2018). Scopus Export 2015-2019. 8789.
https://stars.library.ucf.edu/scopus2015/8789