Title
Z-Scan And P-Scan: New Measurements Of The Nonlinear Refractive Index Of Single Crystal Polymer Pts At 1600 Nm
Abstract
The n2 of PTS was measured with 100-fs pulses at 1600 nm. The value of n2 was 4.0 (±0.5) × 10-4 cm2/GW. It was also demonstrated that another measurement technique, power-scan, gives comparable accuracy to Z-scan but is faster to implement.
Publication Date
1-1-2000
Publication Title
Pacific Rim Conference on Lasers and Electro-Optics, CLEO - Technical Digest
Number of Pages
150-151
Document Type
Article
Personal Identifier
scopus
DOI Link
https://doi.org/10.1109/cleo.2000.906843
Copyright Status
Unknown
Socpus ID
0034546269 (Scopus)
Source API URL
https://api.elsevier.com/content/abstract/scopus_id/0034546269
STARS Citation
Yoshino, Fumiyo; Polyakov, Sergey; and Liu, Mingguo, "Z-Scan And P-Scan: New Measurements Of The Nonlinear Refractive Index Of Single Crystal Polymer Pts At 1600 Nm" (2000). Scopus Export 2000s. 1052.
https://stars.library.ucf.edu/scopus2000/1052